Optimization of Gas-Liquid Chromatographic Parameters for the Multiresidue Analysis of 24 Pesticides

잔류농약 24성분의 다성분 동시분석을 위한 기체크로마토그래피 조건의 최적화

  • 이은주 (부산지방식품의약품안전청) ;
  • 김우성 (부산지방식품의약품안전청) ;
  • 박건상 (부산지방식품의약품안전청) ;
  • 오재호 (부산지방식품의약품안전청) ;
  • 김대병 (부산지방식품의약품안전청)
  • Published : 2000.06.30

Abstract

Optimum parameters were investigated for the simultanious analysis of 24 pesticide residuces using gas-liquid chromatography with electron capture detection. Electronic pressure control(EPC) on column enhanced resolution of 24 analyzes. Using DB-17, SPB-608, and Ultra-2 capillary column without EPC incomplete separation was observed in some pairs of pesticides. When EPC function was adopted, no severe overlapping was observed on SPB-608 column in every pesticides except vinclozolin/acetochlor pair. Total running time was 45 min, much shorter than $69{\sim}81$ min when used without EPC. Limit of determination of each analyze ranged $0.1{\sim}12.9$ ng/mL.

분석 대상인 24종 농약을 극성이 다른 캐필러리 칼럼인 DB-17, SPB-608, 및 Ultra-2를 이용하여 전자적압력조절장치(Electronic pressure control, EPC)가 장착된 HP 5890 series II Plus 가스크로마토그래피에서 전자포획검출기로 동시분석법의 최적조건을 모색하였다. 본 실험에서 EPC 시스템의 압력조절기능을 도입하므로써 다음과 같은 분석향상을 얻었다. DB-17, SPB-608, 및 Ultra-2 캐필러리 칼럼에서 EPC 기능없이 동시 분석한 결과 일부 농약들이 분리되지 않았다. 그러나 EPC 시스템의 압력조절 기능을 적용한 결과 vinclozolin과 acetochlor를 제외한 22종 농약 모두 SPB-608 캐필러리 칼럼에서 동시 분석되었다. 또한 농약들의 분리 시간에도 영향을 주어 총 분석시간이 단축되었다. 즉 EPC 시스템 없이 분석한 시간이 $61{\sim}81$분인데 반해 EPC 시스템도입으로 45분으로 단축되었다. 각 농약들에 대한 검출한계는 $0.1{\sim}12.9$ ng/mL 검출되었다.