Left Ventricular Ejection Fraction Determined by Gated Tl-201 Perfusion SPECT and Quantitative Software
게이트 Tl-201 관류 SPECT와 Cedars 소프트웨어를 이용하여 측정한 좌심실 구혈률
- Hyun, In-Young (Departments of Nuclear Medicine, Inha University College of Medicine) ;
- Kim, Sung-Eun (Departments of Nuclear Medicine, Inha University College of Medicine) ;
- Seo, Jeong-Kee (Departments of Internal Medicine, Inha University College of Medicine) ;
- Hong, Eui-Soo (Departments of Internal Medicine, Inha University College of Medicine) ;
- Kwan, Jun (Departments of Internal Medicine, Inha University College of Medicine) ;
- Park, Keum-Soo (Departments of Internal Medicine, Inha University College of Medicine) ;
- Lee, Woo-Hyung (Departments of Internal Medicine, Inha University College of Medicine)
- 현인영 (인하대학교 의과대학 핵의학과) ;
- 김성은 (인하대학교 의과대학 핵의학과) ;
- 서정기 (인하대학교 의과대학 심장내과) ;
- 홍의수 (인하대학교 의과대학 심장내과) ;
- 권준 (인하대학교 의과대학 심장내과) ;
- 박금수 (인하대학교 의과대학 심장내과) ;
- 이우형 (인하대학교 의과대학 심장내과)
- Published : 2000.06.30
Abstract
Purpose: We compared estimates of ejection fraction (EF) determined by gated Tl-201 perfusion SPECT (g-Tl-SPECT) with those by gated blood pool (GBP) scan. Materials and Methods: Eighteen subjects underwent g-Tl-SPECT and GBP scan. After reconstruction of g-Tl-SPECT, we measured EF with Cedars software. The comparison of the EF with g-Tl-SPECT and GBP scan was assessed by correlation analysis and Bland Altman plot. Results: The estimates of EF were significantly different (p<0.05) with g-Tl-SPECT (
목적: 게이트 Tl-201 관류 SPECT와 Cedars 소프트웨어로 구혈률을 측정하고 게이트 심장 혈액풀 스캔으로 얻은 구혈률과 비교하여 상관과 일치도를 분석하였다. 대상 및 방법: 총 18명을 대상으로 게이트 Tl-201 관류 SPECT와 게이트 심장 혈액풀 스캔을 촬영하였다. 게이트 관류 SPECT를 재구성하고 Cedars 소프트웨어로 구혈률을 측정하였다. 게이트 Tl-201 관류 SPECT와 게이트 심장 혈액풀 스캔으로 얻은 구혈률의 상관을 분석하고 Bland Altman 도표를 그려 일시도를 평가하였다. 결과: 게이트 심장 혈액풀 스캔(