Determination of the Cr2O7 Ion by a New Absorption Spectrometric Method Coupled with a Technique of Flow Injection Analysis

흐름주입분석기법을 접목한 새로운 흡수분광분석법에 의한 Cr2O7 이온의 정량

  • Hwang, Hoon (Department of Chemistry, Kangwon National University)
  • 황훈 (강원대학교 자연과학대학 화학과)
  • Published : 20000800

Abstract

A new absorption spectrometric method for the determination of the Cr$_2O_7^{2-}$ ionin acidic media has been developed. The new method is based on the oxidation-reduction reaction of the HCr$O_4^-$ ion with H$_2$O$_2$forming a deep blue CrO(O$_2$) $_2$ andis coupled with a technique of flow injection analysis(FIA). The new method provides a linear calibration curve which accurately follows the Beer's law over a wide range of the analytical concentrations(2.0 ${\times}$ $10^{-6}$M~8.0 ${\times}$ $10^{-3}$M) of K$_2$Cr$_2$O$_2$. The sensitivity of the new method is approximately two times greater than the current method and the effects of the interfering substances such as V, Co, Ni, Fe, and Mn are almost negligible except Cu.

산성 $K_2Cr_2O_7$ 용액 자체의 흡광도를 측정하는 기존의 분석법과는 달리 산성 $K_2Cr_2O_7$ 용액 속에 존재하는 $HCrO_4^-$ 이온과 과산화수소($H_2O_2$)의 산화-원반응의생성물인 진한 푸른색의 $CrO(O_2)_2$의 흡광도를 측정하는 분석법에 흐름주입분석기법을 접목한 새로운 흡구분광분석법을 개발하였다. 새로운 흡수분광분석법은 넓은 분석농도범위($2.0 {\times}10^{-6}$M~$8.0 {\times}10^{-3}$M)에 걸쳐 Beer의 법칙을 정확히 따르는 $K_2Cr_2O_7$용액의 검정선의 작성이 가능하였다. 새로운 흡수분광분석법은 기존의 흡수분광법에 비해 약 2.2배 가량 더 높은 감도을 보이며, Cu를 제외한 V, Co, Ni,Fe, 그리고 Mn가 같은 물질들에 의한 간섭현상도 거의 무시할 수 있다는 장점을 가진다.