Progress in Medical Physics (한국의학물리학회지:의학물리)
- Volume 11 Issue 2
- /
- Pages.101-107
- /
- 2000
- /
- 2508-4445(pISSN)
- /
- 2508-4453(eISSN)
A study on non-invasive SaO$_2$ measurement algorithm to improve on effect of the motion artifact
동잡음의 영향을 개선한 비관혈식 산소포화도 측정 알고리즘에 대한 연구
Abstract
Pulse oximetry is a non-invasive optical method which measures arterial oxygen saturation with two different wavelength. We can obtain the pulsating component of the arterial blood superimposed on DC level attenuated by venous blood, skin and other nonpulsatile components. This study is based on computing algorithm of oxygen saturation using the integral ratio of pulsatile components. In this algorithm, we used the half cycle of the pulsatile signal rely on arterial contraction. It's period is about 1/4 in 1 cycle. In the result, Our algorithm with 1/4 period of 1 cycle is similar to existing model. Because of removal that A part have low amplitude and possession in long term from calculating, the effect of the motion-artifact is decrease.
펄스옥시메트리는 서로 다른 두 파장의 빛을 이용하여 동맥혈의 산소포화도를 측정하는 비침습성의 광에 의한 측정 방법이다. 동맥혈의 맥동성분은 정맥혈, 피부 그리고 기타의 비맥동성분들에 의해 감쇠되는 직류 성분에 중첩되어 있다. 본 논문에서는 맥동 성분의 적분비를 이용하여 산소포화도를 계산하는 알고리즘을 사용하였다. 알고리즘에서는, 심장의 수축과 관련되는 맥동신호의 1/2 싸이클만을 이용한다. 이용하는 시간은 1 cycle의 약 1/4이다. 1/4 주기를 이용하는 본 알고리즘은 기존의 방식과 유사한 결과가 나타났다. 신호의 진폭이 낮고, 시간의 점유가 많은 부분을 계산에서 배제했기 때문에, 동잡음의 영향이 감소되었다.
Keywords