Abstract
This paper presents the characteristics of the dynamic response and calibration technique on a linear variable differential transformer(LVDT). The linear error of the LVDT was proven $\pm$1% in the static calibration and $\pm$0.5% in the dynamic calibration. In this paper, the linearity error generated in the static and dynamic state of the core movement can be eliminated using the correction algorithem of the static and dynamic state derived from the least square linear approximation for the nonlinearity of the curves of direct data fitting and Lagrange polynomials. With the static and dynamic calibration method, the calibration accuracy of the LVDT can be reduced to within $\pm{0.5%.}$.