A STUDY ON THE HEAT GENERATION OF BONE DRILLING BURS ACCORDING TO IRRIGATION DURING IMPLANTATION OF DENIAL IMPLANT

치과 임플랜트 식립시 관주에 따른 골 천공 기구의 열 발생에 관한 연구

  • Moon, Eun-Soo (Department of Prosthodontics, College of Dentistry, Dankook University) ;
  • Lim, Heun-Song (Department of Prosthodontics, College of Dentistry, Dankook University) ;
  • Cho, In-Ho (Department of Prosthodontics, College of Dentistry, Dankook University)
  • 문은수 (단국대학교 치과대학 치과보철학교실) ;
  • 임헌송 (단국대학교 치과대학 치과보철학교실) ;
  • 조인호 (단국대학교 치과대학 치과보철학교실)
  • Published : 1999.08.01

Abstract

There are a lot of elements affecting the success of implant prosthesis. The quality of surgical procedure is considered as one of the key factors. To avoid the excess heat generation is an important element of successful osseointegration and it can be achieved by using a gentle surgical technique with a sharp instrument in bone drilling. This study was performed to measure and analyze comparatively the heat transmitted to sur-rounding bone at a distance of 0.5mm from the periphery of the drill hole in each drilling stage. The results were as follows. In standard system, the temperature of surrounding bone tissue ranged from $29.2^{\circ}C\;to\;48.3^{\circ}C$ with irrigation and from $34.6^{\circ}C\;to\;84.3^{\circ}C$ without irrigation. And in wide system, the temperature of surrounding bone tissue ranged from $29.5^{\circ}C\;to\;52.5^{\circ}C$ with irrigation and from $34.8^{\circ}C\;to\;87.8^{\circ}C$ without irrigation. And the temperature ranges exceeded the threshold without irrigation, while showing less than the threshold by the cooling effect of irrigation. In comparing standard system with wide system, although there was no significant difference, ${\phi}4.3mm$ pilot and ${\phi}4.3mm$ twist drill of wide system showed high value and wide system showed slightly high elevation of temperature in all depth in fixture installation. In the finite element analysis, the calculated value by the Fourier's cooling law were applied to the bone drilling surface. And through analysis using different irrigation temperatures at $28^{\circ}C,\;15^{\circ}C\;and\;5^{\circ}C$, and according to the time. The result was that the cooling water at least below $15^{\circ}C$ was required to maintain the temperature of surrounding bone less than threshold in bone drilling, the cooling water below $5^{\circ}C$ was required to gain more sufficient cooling effect, and cooling over 5 seconds was needed after bone drilling for sufficient effect.

Keywords