QA of a stereotactic radiosurgery system for clinical application
정위방사선수술 시스템의 임상 적용을 위한 QA
Abstract
We developed a sterotactic radiosurgery system which is comprised of 1) collimators with small circular aperture, 2) an angiographic target localizer, 3) a target localizer used for alignment of planned target position with isocenter of treatment machine, and 4) a treatment planning system named LinaPel. In this study, we performed a series of treatment simulations to specify and analyze geometrical errors contained our in-house radiosurgery system. As results, 1) using Geometrical Phantom(Radionics,USA), the accuracy of target localization by LinaPel was determined as Avg. =(equation omitted) the accuracy of mechanical isocenter was found out to be 0.6
BRW 정위프레임을 사용하여 정위방사선수술을 시행하기 위해 1) 정위방사선수술용 원형 콜리메이터, 2) 혈관조영술을 통하여 치료 위치 결정을 위한 localizer, 3) 결정된 치료위치를 치료기에서 조준하기 위한 target localizer, 4) 정위방사선수술용 컴퓨터 치료계획 시스템으로 구성되는 정위방사 선수술 시스템을 개발하였다. 본 연구에서 저자들은 자체 개발한 정위방사선수술 시스템의 임상 적용에 앞서, 인체 모형 팬톰 및 QA용 팬톰을 사용하여 일련의 모의 치료를 시행하여 봄으로써 위치 오차의 정도를 평가하고자 하였다. 실험결과 ,1) 기하 팬톰을 이용한 컴퓨터 치료계획 시스템의 위치 확인 정확도는 평균값, Avg.=(equation omitted)이 1.0mm 이었으며, 2) 치료기의 중심점의 기계적 정밀도는 0.6