The comparisons of three scatter correction methods using Monte Carlo simulation
몬테카를로 시뮬레이션을 이용한 산란보정 방법들에 대한 비교
Abstract
Scatter correction for single photon emission computed tomography (SPECT) plays an important role to improve image quality and quantitation. The purpose of this study was to investigate three scatter correction methods using Monte Carlo simulation. Point source and Jaszack phantom filled with Tc-99m were simulated by Monte Carlo code, SIMIND. For scatter correction, we applied three methods, Compton window (CW) method, triple window (TW) method, and dual photopeak window (DPW) method. Point sources located at various depths along the center line within a 20-cm phantom were simulated to calculate the window ratios and corresponding scatter fractions by evaluating the polynomial coefficients for DPW method. Energy windows were located in W
단일광자방출촬영 (SPECT) 에 대한 산란보정은 영상질을 개선하고 영상을 정량분석하는데 중요한 역할을 한다. 본 연구의 목적은 몬테카를로 시뮬레이션을 이용하여 에너지 창을 이용한 세 가지 산란보정들에 대해 조사하는 것이었다. 몬테카를로 방법을 이용하여 Tc-99m에서 방출되는 광자의 history를 발생시키는 점선원과 Jaszack 팬텀에 대해 시뮬레이션하였다. 시뮬레이션을 위한 SPECT 시스템의 섬광체 NaI(T1) 는 두께 0.95 cm, 40