Abstract
Total reflection X-ray fluorescence spectroscopy using synchrotron radiation source called as TRSFA was explored to achieve high sensitivities to impurity metals on Si wafer surface. It consists of monochromating part to select a specific wavelength, slit part to shield direct beam and to control monochromated beam, and main chamber to dectect fluorescent X-ray counts of impurities on si wafer. Monochromated X-ray of 10.90 KeV was selected and the optimum total reflection condition on silicon wafer was obtained through tuning the dead time and fluorescent X-ray count of Si and Fe. TRSFA system could increase the sensitivity as high as 50 times in comparision with TRXFA using normal X-ray source. But the trend was varied since the surface conditions of Si wafers and, therefore, the reflectivities were different. Furthemore, there seems to be a promising path to reaching a detection limit useful to the next generation metal impurities control, because Fe impurity below to the $5\times10^{9}\textrm{atomas/cm}^2$ can be detectable through the developed TRSFA system.
방사광을 이용한 전반사 형광 분석법으로 Si 웨이퍼 표면 금속 불순물의 검출능을 향상시켰다. 측정장치는 특정 단색광 에너지만을 선택할 수 있는 모노크로메팅부, 측정챔버안으로 유입되는 방사광은 차폐하고 원하는 크기의 단색광을 선택하는 슬릿부 그리고 Si 웨이퍼 표면에서 전반사에 의해 발생하는 형광 X-선을 검출하는 측정부로 구성되어 있다. 단색광의 에너지는 10.90 KeV로 선택하였고, 최적의 전반사 조건을 확립하기 위하여 소멸시간과 Fe의 형광 X-선의 강도비의 관계를 이용하였다. 기존 X-선원을 이용하여 관찰한 결과와 비교하였을 경우에, 최대 약 50배까지 검출감도를 향상 시킬 수 있었다. 특히, TRXFA(Total Reflection X-ray Fluorescence Analyzer)법으로는 검출하기 어려운 $5\times10^{10}\textrm{atomas/cm}^2$ 수준의 금속오염은 방사광을 이용한 TRSFA(Total Reflection Synchrotron Fluorescence Analyzer)법으로는 충분히 검출할 수 있고, $5\times10^{9}\textrm{atomas/cm}^2$금속 불순물가지 검출할 수 있는 방법 및 장치를 개발하였다. 이를 이용하여 차세대 Si 웨이퍼의 초극미량 금속 불순물 분석에 이용할 수 있는 방법으로 기대된다.