Abstract
Laboratory experiment was performed for the TDR to measure the soil moisture, and the results, were compared with the design water content and the one measured by oven-try method. Sand and kaolinite were used . Varaiables for the experiment were water content (10-50%), void ration (0.7 -1.3), mixture rate of kaolinite (10-30%), and measurement methods (TDR and oven-dry). In all cases , TDR method showed very accurage and reliable results , and average error and error range were far lews than the oven-dry method which is widely used. Considerable error was noticed when water contnet was 50% where saturation was achieved for both methods. Therefore, TDR was thought to be applicable to the field moisture measurement if it is unsaturated. For field scale application of TDR, more research and verification of the accuracy with diverse soil conditions including physical ,chemical and mineral properties are recommended.