Abstract
As a part of trying to get the resistivity values correctly from laboratory core resistivity measurement, the effect of sample holders in resistivity measurement was analyzed and a better way to determine the representative resistivity value from the time series resistivity data was searched. Modified GS type and modified two-electrode type sample holders were devised and their effects have been compared with those of GS and two-electrode type sample holders. The modified two-electrode type sample holder has benefits both in repetition and simplicity in data acquisition. The analysis of distribution trend of the time series resistivity data obtained with different kind of sample holders and source frequencies shows that the maximum curvature point method gives the best result in determining representative resistivity value.
암석의 코어비저항 값을 올바르게 결정하기 위한 노력의 일환으로 비저항 측정에 있어서 샘플홀더의 영향을 분석하고 시계열자료로부터 암석의 비저항을 대표할 수 있는 비저항 값을 효과적으로 구하는 방안에 관하여 연구하였다. 샘플홀더에 대한 연구로는 기존의 GS식과 2전극식 방법과 함께 코어시료 표면을 전도성 접착제로 처리한 변형 GS식 및 변형 2전극식을 고안하여 각 특성을 비교하였으며 그 중에서 변형 2전극식이 측정자료의 안정성과 측정의 편이성 측면에서 장점이 있는 것으로 확인되었다. 샘플홀더 및 사용 소스 주파수의 차이에 따라 달리 나타나는 시계열자료의 분포특성에 관한 분석 결과는 최대곡률점을 이용하여 암석의 대표비저항을 결정하는 방법이 효과적임을 보이고 있다.