The useage of the EPID as a QA tools

EPID의 적정관리 도구로서의 유용성에 관한 연구

  • Cho Jung Hee (Department of Radiation Oncology, Yonsei Cancer Center) ;
  • Bang Dong Wan (Department of Radiation Oncology, Yonsei Cancer Center) ;
  • Yoon Seong Ik (Department of Radiation Oncology, Yonsei Cancer Center) ;
  • Park Jae Il (Department of Radiation Oncology, Yonsei Cancer Center)
  • 조정희 (연세 암센터 방사선종양학과) ;
  • 방동완 (연세 암센터 방사선종양학과) ;
  • 윤성익 (연세 암센터 방사선종양학과) ;
  • 박재일 (연세 암센터 방사선종양학과)
  • Published : 1999.12.01

Abstract

Purpose : The aim of this study is to conform the possibility of the liquid type EPID as a QC tools to clinical indication and of replacement of the film dosimetry. Aditional aim is to describe a procedure for the use of a EPID as a physics calibration tool in the measurements of radiation beam parameters which are typically carried out with film. Method & Materials : In this study we used the Clinac 2100c/d with EPID. This system contains 65536 liquid-filled ion chambers arranged in a $256{\times}256$ matrix and the imaging area is $32.5{\times}32.5cm$ with liquid layer thickness of 1mm. The EPID was tested for different field sizes under typical clinical conditions and pixel values were calibrated against dose by producing images using various thickness of lead attenuators(lead step wedge) using 6 & 10MV x-ray. We placed various thickness of lead on the table of linear accelerator and set the portal vision an SDD of 100cm. To acquire portal image we change the field size and energy, and we recorded the average pixel value in a $3{\times}3$ pixel region of interest(ROI) at field center was recorded. The pixel values were also measured for different field sizes in order to evaluate the dependence of pixel value on x-ray energy spectrum and various scatter components. Result : The EPID, as a whole, was useful as a QA tool and dosimetry device. In mechanical check, cross-hair centering was well matched and the error was less than ?2mm and light/radiation field coincidence was less than 1mm also. In portal dosimetry the wider the field size the the higher the pixel value and as the lead thickness increase, the pixel value was exponentially decreased. Conclusions : The EPID was very suitable for QA tools and it can be used to measure exit dose during patients treatment with reasonable accuracy. But when indicate the EPID to clincal study deep consideration required

Keywords