선형가속기에서 상위조리개와 하위조리개의 교환에 의한 선량 변화의 고찰
The study on dose variation due to exchange of Upper and Lower jaw in the linear accelerator
초록
The field size can be beam output, therefore MonitorUnit can be varied due to field size dependence The purpose of this study is to evaluate and compare the dose variation according to exchange of collimator The measurements were perfomed with Wellhofer dosimetry system(water phantom. ion chamber. electrometer. system controller. build up cap. etc)and two types of linear accerlerator (Mevatron KD, MevatronMX) Scatter can be affected to field size dependence and scatter correction is separated into collimator and phantom components, scatter components can affect by exchanging of collimator Measurements of collimator scatter factor(Sc) was done in air with build up cap. 1)Square field (5cm2 to 40cm2) was measured 2)and then keeping the upper jaw constant at loom and varing lower jaw from 5cm to 40cm, 3)keeping the lower jaw constant at 10cm and varing upper jaw from 5cm to 40cm Measurements of total scatter factor(Scp) was done in water at Dmax as the procedure of collimator scatter factor measurements in water Dmax The total scatter factors were obtained to the following equation(Sp=Scp/Sc) The measured data is normalized to the data of reference field size(