Acknowledgement
Supported by : 삼성의료원
Recently electronic apex locators have been used widely in root canal treatment, but, accuracy of electronic apex locators is controversial. The purpose of this study was to evaluate the accuracy of Apex Finder A.F.A(EIE Analytic Technology, U.S.A.) in vivo compared with Root-Zx and radiograph. The root canal lengths were determined with Root-Zx(32 tooth) in before pulp extirpation and after pulp extirpation. Then the radiographs were taken with a file in the canal. The root canal lengths were determined with Apex Finder A.F.A.(21 tooth) in before pulp extirpation and after pulp extirpation and under NaOCl. Then the radiographs were taken with a file in the canal. The results were as follows: 1. There was no significant statistical difference in Root-Zx between before pulp extirpation and after pulp extirpation(p > 0.05). 2. There was no significant statistical difference in Apex Finder A.F.A. between before pulp extirpation and after pulp extirpation(p > 0.05). But, there was significant statistical difference under NaOCl(p < 0.05). 3. There was no significant statistical difference in accuracy between Root-Zx and Apex Finder A.F.A.
Supported by : 삼성의료원