Abstract
A simple on-line measurement system consisting of a conventional peristaltic pump, a HPLC-type heater, and a flow-through spectrophotometer is introduced for the determination of chemical oxygen demand(CODI. The system was configured such that the reaction mixture in the highly concentrated surffuric acrid medium flowing through the PTFE reaction tubing was heated at 150℃ and the absorbance of dichromate was continuously moutored at 445 m. The same addation principle as in the standard procedure was employed akcept the use of CoSO4 as a new effective catalyst. To test the system, potassium hydrogen phthalate was selected as a COD standard material. With suitably optimized reaction condition, the applicable concentration range depends on the concentration of potassium dichromate in the oxidizing reagent. With 2.0×10-3 M and 5.0×10-4M dichromate, the linear dynamic range was observed up to 400 ppm and 100 ppm, respectively. The standards in the Unear ranges were shown to be completely oxidized, which was confirmed with sodium oxalate or Mohr's salt. In all cases, the typical reproduclbility for betweenruns was 2% or less. The proposed measurement system provides the valuable in- formation for the further development of automated analysis system based on the present standard procedure.