Abstract

A portable HTS RF SQUID-based system, weighing less than 20 kg has been built for susceptometry applications in weak magnetic fields, It includes a YBCO sensor for measuring the axial magnetic field component with a resolution of about $7{\times}10^{-13} T/Hz^{1/2}.$ This is determined by the intrinsic magnetic noise in the quasi-white noise region. There is a relaxation for a sudden increase in field due to magnetic flux creep in HTS. In this instance the time did not exceed 3~5 minutes.

Keywords

References

  1. J. of Low Temp. Phys. v.76 no.5;6 T. Ryhanen;H. Seppa
  2. Rev. Sci. Instrum. v.48 J. S. Philo;W. M. Fairbank
  3. IEEE Trans. Mag. v.28 R. E. Sager;A. D. Hibbs;S. Kumar
  4. Cryogenics v.33 no.6 D. A. Konotop;V. I. Shnyrkov;V. P. Timofeev;S. S. Khvostov;G. M. Tsoi
  5. IEEE Trans. Appl. Supercon. v.7 no.2 J. W. Purpura;R. F. Wieger