The Suppressing of MR Image Artifacts using Phases Cycling in Fast SE Sequence

  • Published : 1998.06.01

Abstract

The correction of image artifacts due to misadjustment in tuning of RF coils (tip angle) and in the RF single sideband spectrometer was investigated using phase cycling of the $\pi$/2 and $\pi$ pulses in spin-echo sequences. A general procedure was developed for the systematic design of phase cycles that select desirable coherence transfer pathways. To analyze a phase cycling sequence, changes in the coherence level and phase factor for each RF pulse in the spin-echo cycle must be determined. Four different phase cycling schemes (FIXED, ALTERNATE, FORWARD, REVERSED) to suppress unwanted signal components such as mirror and ghost images were evaluated using two signal acquisitions. When the receiver phase factor is cycled counter-clockwise (REVERSED), these artifacts are completely removed.