Development and Performance of Automated Calibration System of Sound Level Meters

소음계 교정 자동화 시스템 개발 및 성능평가

  • 김용태 (한국표준과학연구원 음향진동그룹) ;
  • 조문재 (한국표준과학연구원 음향진동그룹) ;
  • 이용봉 (한국표준과학연구원 음향진동그룹) ;
  • 서재갑 (한국표준과학연구원 음향진동그룹)
  • Published : 1998.10.01

Abstract

An automated calibration system of sound level meters was developed and tested. As a standard sound source, the speaker unit(Forstex FE208) cabineted by 440$\times$390$\times$490 $\textrm{mm}^3$(LHW) volume wood box was adopted. Including this source, the driving part was found out to have a good linearity of sound pressure output vs AC voltage input. The Hybrid-Bisect/Newton-Raphson method modified by the linearity was adopted as a searching algorithm. Uisng GPIB interface, the console PC make the control, measurements, and calculations and finally make the accumulation of useful data and results automatically by the instructon in the program coded by C languate. Several trials of automatic calibration using this developed system give the reliable results.