The Broad-beam CT Image Reconstruction from Simulator Images

모의치료(Simulation) 영상을 이용한 Broad-beam CT 영상 구현

  • Yi, Byong-Yong (Department of Radiation Oncology, Asan Medical Center, College of Medicine University of Ulsan)
  • 이병용 (울산대학교 의과대학 서울중앙병원 방사선종양학교실)
  • Published : 1998.03.01

Abstract

Purpose : To generate the axial, coronal and sagittal images from conventional simulation images, as a preliminary study of broad-beam simulator CT. Methods and Materials : Volumetric filtered back-projection was performed using 90 sheets of films from conventional simulator for every $4^{\circ}$ gantry angle. Two mAs exposure condition for 120kvp beam qualify at SFD 140cm was given to each film. Outside the silhouette portion was removed and scatter component was deconvolved before back-projection. Results : The axial, the sagittal and the coronal images with same spatial resolutions over all direction could be obtained. But image quality was very poor. Conclusion : CT images could be obtained using broad-beam. Scatter deconvolution technique was effective for this reconstruction. The fact that same spatial resolutions over all direction tells us the possibility of application of this technique to DRR or Simulator-CT. But the quality of image should be improved for clinical application practically.

목적 : Broad-beam Simulator-CT 개발을 위한 예비연구로서 통상의 모의 치료 영상에서 축상면, 시상면, 관상면 영상을 구현하려 하였다. 대상 및 방법 : 120kVp, 2mAs 동일 조건에서 갠트리 각도를 $4^{\circ}$ 간격으로 90장의 필름을 얻어 입체적인 Filtered back-projection을 시도하였다 외곽선을 찾아 제거하였고, 산란선 성분을 Deconvolution 방법으로 제거하여 좋은 영상을 얻도록 하였다. 결과 : 이 방법으로 축상면, 시상면, 관상면 영상을 얻었으며 각 방향에 대해 동일한 분해능을 갖았다. 그러나 영상의 질은 대단히 나빴다. 결론 : Broad-beam으로 된 CT 영상을 구현할 수 있었다. 이를 위하여 산란선 성분의 Deconvolution이 필요하였으며, 입체적인 back-projection을 실시하였으므로 축상, 시상, 관상 모든 방향에 대해 동일한 분해능을 갖고 있어서 DRR 등 Simulator-CT에 응용할 수 있음을 알 수 있었다. 그러나 실용적인 임상응용을 위해서는 영상의 질 개선이 필요하였다.

Keywords