Intercomparison Study of the Neutron Personnel Dosemeters

중성자 개인선량계 상호비교

  • Published : 1998.03.30

Abstract

Domestic intercomparison study of the neutron personnel dosemeters was performed for the first time in Korea. Thirteen types of neutron dosemeters from twelve institutions took part in this intercomparison study and the $D_2O$ moderated Cf-252 source of KAERI was used for irradiation. Eight of the fifteen dosemeters submitted by each participant were divided into two groups and each group was irradiated with different doses of the simulated mixed fields of neutron and gamma. The participants assessed their dosemeter reading in terms of the personal dose equivalent, Hp(10), for both neutron and gamma dose. The ratio of the reported dose equivalent to the delivered dose equivalent for comparison between participants ranged from 0.55 to 1.34 for neutron, from 0.54 to 1.32 for gamma and from 0.75 to 1.20 for total dose. This intercomparison results show that all dosemeter processors, especially for neutron category, are able to pass the personnel dosemeter performance test which shall be enforced according to the ordinance of the MOST, No. 96-6.

국내 최초로 중성자 개인선량계에 대한 상호비교측정시험이 수행되었다. 기준 방사선장으로 한국원자력연구소가 보유하고 있는 중수감속 $^{252}Cf$ 선원을 이용하였으며, 12개 판독기관의 선량계 13종이 상호비교시험에 참가하였다. 각 참가기관으로부터 컨트롤과 예비용을 포함하여 15개의 선량계를 제출받아, 이를 2개의 조사선량군으로 나누어 4개씩 총 8개의 선량계가 실제 조사되었다. 중성자, 감마 그리고 총선량의 항목으로 판독기관의 보고선량을 부여된 선량으로 나누어 선량계 판독결과를 비교한 결과, 각각에 대하여 그 비율이 $0.55{\sim}1.34$, $0.54{\sim}1.32$, $0.75{\sim}1.20$ 의 분포를 갖는 것으로 나타났다. 판독기관의 자체 판독능력을 기준으로 할 때 전혀 문제가 없는 것은 아니나, 현재의 상호비교시험 결과로부터 알 수 있는 것은 향후 중성자분야에 대한 개인선량계 성능시험이 시행될 경우, 판독기관들이 모두 합격범위에 들 가능성이 높은 것으로 평가되었다.