Establishment of the Monoenergetic Fluorescent X-ray Radiation Fields
교정용 단일에너지 형광 X-선장의 제작
- Kim, Jang-Lyul (Korea Atomic Energy Research Institute) ;
- Kim, Bong-Hwan (Korea Atomic Energy Research Institute) ;
- Chang, Si-Young (Korea Atomic Energy Research Institute) ;
- Lee, Jae-Ki (Han-Yang University)
- Published : 1998.03.30
Abstract
Using a combination of an X-ray generator Installed in radiation calibration laboratory of Korea Atomic Energy Research Institute (KAERI) and a series of 8 radiators and filters described in ISO-4037, monoenergetic fluorescent X-rays from 8.6 keV to 75 keV were produced. This fluorescent X-rays generated by primary X-rays from radiator were discriminated
한국원자력연구소 교정시설에 설치되어 있는 MG325 X-선 발생장치와 ISO-4037에서 제시하고 있는 라디에이티 및 필터 8종을 조합하여 8.6 keV 부터 75 keV 까지의 단일에너지 형광 X-선을 제작하였다. 1차 X-선에 의하여 라디에이터에서 발생된 형광 X-선중