An Evaluation of the Accuracy of Mini-Wright Peak Flow Meter
mini-Wright Peak Flow Meter에 의한 PEFR 측정의 정확도
- Koh, Young-Il (Department of Internal Medicine, and Research Institute of Medical Science Chonnam University Medical School) ;
- Choi, In-Seon (Department of Internal Medicine, and Research Institute of Medical Science Chonnam University Medical School) ;
- Na, Hyun-Ju (Department of Internal Medicine, and Research Institute of Medical Science Chonnam University Medical School) ;
- Park, Seok-Chae (Department of Internal Medicine, and Research Institute of Medical Science Chonnam University Medical School) ;
- Jang, An-Soo (Department of Internal Medicine, and Research Institute of Medical Science Chonnam University Medical School)
- 고영일 (전남대학교 의과대학 내과학교실 및 의과학 연구소) ;
- 최인선 (전남대학교 의과대학 내과학교실 및 의과학 연구소) ;
- 나현주 (전남대학교 의과대학 내과학교실 및 의과학 연구소) ;
- 박석채 (전남대학교 의과대학 내과학교실 및 의과학 연구소) ;
- 장안수 (전남대학교 의과대학 내과학교실 및 의과학 연구소)
- Published : 1997.04.30
Abstract
Background : Portable devices for measuring peak expiratory flow(PEF) are now of proved value in the diagnosis and management of asthma and many lightweight PEF meters have become available. However, it is necessary to determine whether peak expiratory flow rate(PEFR) measurements measured with peak flowmeters is accurate and reproducible for clinical application. The aim of the present study is to define accuracy, agreement, and precision of mini-Wright peak flow meter(MPFM) against standard pneumotachygraph. Methods : The lung function tests by standard pneumotachygraph and PEFR measurement by MPFM were performed in a random order for 2 hours in 22 normal and 17 asthmatic subjects and also were performed for 3 successive days in 22 normals. Results : The PEFR measured with MPFM was significantly related to the PEFR and
연구배경 : 휴대용 PFMs에 의한 PEFR측정은 쉽고 간편하여 천식의 진단과 추적관찰에 널리 이용되고 있는데, 이들에 의해 측정된 PEFR성적이 얼마나 믿을만 할 것인지는 의문이다. 따라서 저자는 전산화폐기능검사기계인 pneumotachygraph를 기준으로 MPFM에 의한 PEFR측정의 정확도, 일치성, 정밀성을 검토하고자 하였다. 방 법 : 22례의 정상인과 17례의 경증 및 중등도의 천식환자를 대상으로 휴대용 PFMs인 MPFM로 PEFR을 측정하였고 전산화폐기능검사기계인 pneumotachygraph로 폐환기기능검사를 시행하였으며, 이중 정상인만을 대상으로 서로 다른 3일 동안 연속적으로 같은 방법으로 폐기능검사를 시행하여 다음과 같은 결과를 얻었다. 결 과 : 정상인과 환자군 모두에서 MPFM로 측정한 PEFR은 pneumotachygraph 에 의해 측정한 PEFR,