A Study on the Analysis of Histamine by Spectrofluorimetry

형광측정법을 이용한 히스타민 분석방법 개발 연구

  • 이재성 (한국과학기술연구원 특성분석센터) ;
  • 김영상 (고려대학교 화학과) ;
  • 차기석 (한국과학기술연구원 특성분석센터) ;
  • 이희정 (고려대학교 화학과) ;
  • 홍천수 (연세대학교 의과대학 내과학교실) ;
  • 남혜원 (수원여자전문대학 식품영양학과) ;
  • 명춘옥 (오산전문대학 식품조리학과) ;
  • 이기완 (경원대학교 식품영양학과) ;
  • 이양자 (연세대학교 식품영양학과)
  • Received : 1996.10.04
  • Published : 1997.02.25

Abstract

The application of high performance liquid chromatogaphy(HPLC) with fluorimetric detection for the measurement of histamine in biological fluids has been developed recently. Because of time consumption during measurement and difficulties in controling optimum operating condition for HPLC system, spectrofluorimetry was adopted to detect histamine. Detection limit by spectrofluorimetry was $3.0{\times}10^{-2}{\mu}g/mL$ wi th R. S. D. 3.74%~16.2% on standard solution. The relative ratio of histmine levels between the result from HPLC-fluorimetry and that from spectrofluorimetry ranged from 1.5 to 3.5. The recoveries of each samples obtained from standard addition method by spctrofluorimetry have shown 100%~125%.

생물학적 분비물에서 히스타민의 측정시 HPLC-fluorimetry가 많이 응용되어 발달되어 왔다. 그러나 시간 소비와 HPLC system을 최적의 운영조건으로 유지시키는 것이 상대적으로 어렵기 때문에 직접 형광법을 이용하여 히스타민을 분석하였다. 표준 용액에서 형광측정법의 경우 검출 한계는 $3.0{\times}10^{-2}{\mu}g/mL$였고 이 때 상대표준편차는 3.74%~16.2%였다. 시료에서 HPLC-fluorimetry와 형광측정법 사이의 상대적인 히스타민 농도비율은 1.5~3.5배의 차이를 보였다. 표준물첨가법에 의한 각각의 시료에 대한 수득률은 100%~125%였다.