Abstract
Zinc, cadmium, lead and copper were simultaneously determined by depositing metals at - 1.200 V vs. a Ag/AgCl(sat. KCl) reference electrode for 150 seconds on a hanging mercury drop electrode(HMDE) or a thin mercury film electrode(TMFE), followed by scanning towards anodic direction using differential pulse voltammetric(DPASV) and square wave voltammetric(SWASV) techniques. The linear calibration curves were obtained for four metal ions simultaneously determined by DPASV with a HMDE in the concentration range between 20 and 100 ppb. However, the linear calibration plots were obtained only for $Cd^{2+}$ and $Pb^{2+}$ in the simultaneous determinations with a TMFE in the concentration range up to 100 ppb using DPASV and up to 10 ppb using SWASV. DPASV with a TMFE was about 15 times more sensitive than DPASV with a HMDE for simultaneous determinations of $Cd^{2+}$ and $Pb^{2+}$. SWASV was about 5 times more sensitive than DPASV at a TMFE. Concentrations of zinc in seven different sediment samples determined by DPASV with a HMDE and inductively coupled plasma-mass spectrometry were compared, resulting with an excellent correlation coefficient of 0.9993 and with no significant difference between two methods after t-test.
매달린 수은 방울 전극(HMDE) 또는 얇은 수은막 전극(TMFE)을 사용하여 금속이온들을 은/염화은(포화 KCl) 기준전극에 대하여 -1,200 V에서 150초 동안 전해시켜서 수은전극에 농축시키고 펄스 차이 전압전류법(DPASV)과 네모파 전압전류법(SWASV)으로 산화전극 벗김 분석을 하여 동시에 아연, 카드뮴, 납 및 구리를 정량분석하였다. HMDE를 사용하여 DPASV로 네 가지 금속이온을 동시 정량분석시 각각의 금속이온의 봉우리 전류는 20~100 ppb 농도범위에서 직선성을 보여주었으나 TMFE를 사용하여 DPASV 또는 SWASV로 네 가지 금속이온을 동시 정량분석시에는 $Cd^{2+}$와 $Pb^{2+}$의 봉우리 전류만 DPASV의 경우 100 ppb까지 SWASV의 경우 10 ppb까지 직선성을 나타내었다. $Cd^{2+}$와 $Pb^{2+}$의 동시 정량분석의 경우 TMFE를 사용한 DPASV 분석은 HMDE를 사용한 DPASV보다 약 15배 더 민감하였으며 TMFE에서 SWASV는 DPASV보다 약 5배 더 민감하였다. 퇴적물에 함유된 아연의 농도를 HMDE를 사용한 DPASV 분석법과 유도 결합 플라스마-질량분석법으로 일곱개의 시료에 대하여 정량분석하여 비교하였더니 상관계수가 0.9993으로 높았고 t-test결과 두 방법 사이에는 유의성 있는 차이가 없었다.