Silicone Rubber Membrane-Based Chloride Ion-Selective Electrode

실리콘러버-지지체 염화이온선택성 막전극

  • Sakong, Dong-Sik (Department of Chemistry and REsearch Institute of Basic Science, Kwangwoon University) ;
  • Kim, Jin-Mog (Department of Chemistry and REsearch Institute of Basic Science, Kwangwoon University) ;
  • Cha, Jin-Mog (Department of Chemistry and REsearch Institute of Basic Science, Kwangwoon University)
  • 사공동식 (광운대학교 이과대학 화학과,기초과학연구소) ;
  • 김진목 (광운대학교 이과대학 화학과,기초과학연구소) ;
  • 차근식 (광운대학교 이과대학 화학과,기초과학연구소)
  • Published : 19971200

Abstract

Silicone rubber-matrix membranes doped with tetradecyltrimethylammonium chloride (TDTMACl) are used to enhance electrode performance for chloride measurements in physiological samples. The optimized membrane formulation incorporates 95.4 wt% silicone rubber and 4.6 wt% TDTMACl, and its pH response is negligible in the range of pH 6-10. The TDTMACl-doped silicone rubber membrane exhibits sub-Nernstian response to chloride from 10 to 300 mM (-37.5 mV/decade), but its selectivities for chloride over other anions are remarkably enhanced: KpotCl,NO3=1.3, KpotCl,I=2.0, KpotCl,Sal=0.8, KpotCl,SCN=2.0 and KpotCl,ClO4=0.8. Furthermore, since the silicone rubber-matrix membrane exhibits better adhesion to the solid surface than do PVC membranes, the lifetime of the coated-wire type membrane electrode is greatly improved. The response properties toward chloride and salicylate for solid-state sensor are not significantly changed after at least 35 days of use.

Tetradecyltrimethylammonium chloride(TDTMACl)을 첨가한 실리콘러버-지지체 전극막은 생체시료중 염화이온의 측정시 보다 향상된 전기화학적 특성을 나타내기 위하여 사용되었다. 이때 최적화된 전극막의 조성비는 95.4 wt% 실리콘러버, 4.6 wt% TDTMACl이며 이 전극막의 pH 감응성은 pH 6-10 영역에서 무시할 수 있었다. 이들 전극막들의 염화이온에 대해 감응기울기는 10-300 mM 영역에서 -3.75 mV/decade로 이론적인 Nernstian 감응기울기에는 못 미쳤으나 살리실레이트 이온 등 다른 음이온에 대한 선택성은 매우 우수하였다: KpotCl,NO3=1.3, KpotCl,I=2.0, KpotCl,Sal=0.8, KpotCl,SCN=2.0, KpotCl,ClO4=0.8. 또한 실리콘러버-지지체 전극막은 PVC-지지체 전국막에 비해 고체상 전극표면에 대한 접착력이 우수하므로 CWE형 고체상 전극에 도입되었을 때 향상된 전극수명을 나타내었다. 이들 고체상 전극들은 35일까지 염화이온에 대한 감응기울기나 살리실레이트 이온에 대한 감응성에 변화없이 우수하게 작동함을 알 수 있었다.

Keywords

References

  1. Principles of Clinical Chemistry Blick, K. E.;Liles, S. M.
  2. Ion-Selective Electrode Methodology v.1;2 Covington, A. K.
  3. Ion-Selective Electrodes in Analytical Chemistry v.1 Freiser, H.
  4. Clin. Chem. v.32 Oesch, U.;Ammann, D.;Simon, W.
  5. Electroanalysis v.3 Lewenstam, A.;Maj-Zurawska, M.;Hulanicki, A.
  6. Biomaterials v.13 Davies, M. L.;Hamilton, C. J.;Murphy, S. M.;Tighe, B. J.
  7. Arch. Exp. Pathol. Pharmakol v.24 Hofmeister, F.
  8. Anal. Chem. v.61 Chaniotakis, N. A.;Park, S. B.;Meyerhoff, M. E.
  9. J. Electroanal. Chem. v.335 Kibbey, C. E.;Park, S. B.;DeAdwyler, G.;Meyerhoff, M. E.
  10. Electroanalysis v.1 Brown, D. V.;Chaniotakis, N. A.;Lee, I. H.;Ma, S. C.;Park, S. B.;Meyerhoff, M. E.;Rick, R. J.;Groves, J. T.
  11. Anal. Sci. v.8 Jyo, A.;Egawa, H.
  12. Anal. Chem. v.60 Chaniotakis, N. A.;Chasser, A. M.;Meyerhoff, M. E.;Groves, J. T.
  13. Electroanalysis v.3 Park, S. B.;Matuszewski, W.;Meyerhoff, M. E.;Liu, Y. H.;Kadish, K. M.
  14. Anal. Chim. Acta v.271 Rothmaier, M.;Simon, W.
  15. Anal. Chim. Acta v.327 Rothmaier, M.;Schaller, U.;Morf, W. E.;Pretsch, E.
  16. Anal. Chim. Acta v.315 Cha, M. J.;Shin, J. H.;Oh, B. K.;Kim, C. Y.;Cha, G. S.;Shin, D. S.;Kim, B.
  17. Sens. Actuators B v.32 Sakong, D. S.;Cha, M. J.;Shin, J. H.;Cha, G. S.;Ryu, M. S.;Hower, R. W.;Brown, R. B.
  18. Anal. Chim. Acta v.180 Thomas, J. D. R.
  19. Anal. Chem. v.53 Oesch, U.;Caras, S.;Janata, J.
  20. Analyst v.113 Moody, G. J.;Thomas, J. D. R.;Slater, J. M.
  21. Anal. Chem. v.63 Cha, G. S.;Liu, D.;Meyerhoff, M. E.;Cantor, H. C.;Midgley, A. R.;Goldberg, H. D.;Brown, R. B.
  22. Sens. Actuators B v.33 Hower, R. W.;Shin, J. H.;Cha, G. S.;Meruva, R. K.;Meyerhoff, M. E.;Brown, R. B.
  23. Anal. Chem. v.68 Shin, J. H.;Sakong, D. S.;Nam, H.;Cha, G. S.
  24. Anal. Chim. Acta v.164 Gadzekpo, V. P. Y.;Christian, G. D.
  25. Methods in Clinical Chemistry Pesce, A. J.;Kaplan, L. A.