Abstract
Calcium isotopic ratios are precisely measured by removing isobaric interferences originated from water in the plasma. Liquid Ar cryogenic trap combined with membrane desolvator could eliminate backgrounds at m/z 42 and 44. Slow drift of ICP-MS is corrected by the frequent running of the standards. It is found necessary to separate Ca from the sample matrix using Ca oxalate precipitation technique. Currently, the RSD is 0.5-1.0% for 2 minutes of measurement but is expected to be improved if the measurement time is increased. The technique was applied to 42Ca enriched baby fecal samples and successfully determined 42Ca/44Ca ratio changes.