Animal Experiment and Blood Biocompatibility Study of Polymer Valve

고분자판막의 동물실험 및 생체내 혈액적합성 연구

  • 김상현 (연세대학교 의과대학 심혈관연구소) ;
  • 홍유선 (연세대학교 의과대학 심혈관연구소, 연세대학교 의과대학 흉부외과학교실)
  • Published : 1997.04.01

Abstract

In vivo testings of the monoleaflet polymer valve were performed in seven dogs to prove its blood biocompatibility. The monoleaflet polymer valve used in this study was developed for short-term usage n the ventricular assist device. The frame and leaflet of the polymer valve were made of polyurethane. The inter-aortic valved conduit were implanted in four dogs and the ventriculo-atrial valved conduit was implanted in one dog. The ventricular assist devices with polymer valve were implanted in two dogs. The longest survival was 20 days. Main causes of death were bleeding and infection. To examine the blood compatibility, each blood sample was collected and RBC, WBC, hematocrit, hemoglobin, platelet and lactic acid dehydrogenase were analyzed. These studies thus far demonstrated that, with further development, a reliable and inexpensive polymer valve will be used in the ventricular assist device as short term usage.

심실보조장치 등의 단기간 사용을 위하고 경제적이고 항혈전성이 높은 판막을 개발하기 위해 고분자 판막을 연구 중인데 판막의 혈액적합성을 시험하기 위해 동물실험을 시행하였다. 고분자판막은 외경이 각각 22 mm, 18 mm, 16 mm인 단엽식 판막이다. 열림판과 프레 임 모두 재질을 폴리우레탄 계로 제작하 여 유연하고 완전한 접착이 되도록 하였다. 총 7 마리의 황견에 이식하였는데 이식 위치에 따라 3가지로 분류하였다. 첫 번째는 심실보조장치의 유입구에 삽입하였고,두 번째는 대동맥에,세 번째는우심실과 우심방사이에 인조혈관을 이용하여 이식하였다. 가장 긴 생존 기간은 20일 이었다. 판막의 항혈전성 및 용혈현상을 입증하기 위 해 헤모글로빈, 헤마토크리트, 적혈구수, 백혈구수, LDH, 혈색소 등을 검사하였 다. 본 연구는 좀 더 개선하면 고분자판막이 심실보조장치에 단기간 사용 가능한 판막이 될 수 있는 가능성을 보여주었다.

Keywords

References

  1. J Appl Physiol v.14 Polyurethane artificial heart valves in animals Akutsu T;Dreyer B;Kolff WJ
  2. ASAIO Trans v.28 A polyurethane trileaflet cardiac valve prosthesis: In vitro and in vivo studies Wisman CB;Pierce WS;Donachy JH;Pac WE;Myers JL;Prophet GA
  3. ASAIO Trans v.35 New Polyurethane valves in new soft artificial hearts Yu LS;Yuan B;Bishop D(et al.)
  4. ASAIO Trans v.V57 Advances in design principle and fluid dynamics of a flexible polymeric heart valve Jansen J;Willeke S;Reiners B(et al.)
  5. ASAIO Trans v.34 A newly designed jellyfish valve for an artificial heart blood pump Imachi K;Fujimasa I;Mabuchi K(et al.)
  6. The Int J of Art Organ v.13 Development of a new inflow valve for a 20cc semisoft ventricle: preliminary results Wijsmuller EG;Yu LS;Yuan B;Bishop ND;Kolff WJ
  7. 의공학회지 v.13 이엽식 고분자판막의 수력학적 특성평가 및 동물실험 김철생;박복춘;서수원;한동근;이규백;최진욱(등)
  8. 의용생체공학회 춘계학술대회논문집 단엽식 고분자판막의 모의순환 실험 김상현;김원기;장병철;조범구
  9. 의공학회지 v.17 응력 및 변위를 최소화하기 위한 단엽식 고분자 판막의 유한 요소 해석 한근조;안성찬;김상현;신정욱
  10. 의공학회지 v.16 단엽식 고분자판막의 혈역학적 성능평가 김상현;장병철;박영환;강면식;조범구;한근조;이종선
  11. 의공학회지 v.14 성인용 공기구동식 심실보조장치의 모의순환실험 김상현;장병철;김원기;김남현;조범구
  12. 대흉회지 v.30 no.3 공압식 심실보조장치의 개발 및 동물실험 박영환;김상현;홍유선;조범구;장병철;강면식
  13. J. Biomechanical Engineering v.114 Numerical simulation of steady flow in a two-dimensional total artificial heart model Kim, SH;Chandran KB;Chen CJ