A new gradient coil design technique for open magnetic resonance imaging systems

개방형 자기공명영상시스템용 경사자계코일의 새로운 설계기법

  • Lee, Soo-Yeol (Dept. of Biomedical Eng., College of Medicine, Kongkuk University) ;
  • Park, Bu-Sik (Dept. of Biomedical Eng., College of Medicine, Kongkuk University) ;
  • Yi, Jeong-Han (Medical Equipment Lab., Samsung Advanced Institute of Technology) ;
  • Yi, Wan (Medical Equipment Lab., Samsung Advanced Institute of Technology)
  • 이수열 (건국대학교 의과대학 의공학과) ;
  • 박부식 (건국대학교 의과대학 의공학과) ;
  • 이정한 (삼성종합연구원 의료기기실) ;
  • 이완 (삼성종합연구원 의료기기실)
  • Published : 1997.01.01

Abstract

Most open magnetic resonance imaging systems have used the planar gradient coils whose inductances were minimized through the magnetic energy minimization procedure in the spatial frequency domain. Though the planar gradient coils have smaller inductance than conventional gradient coils, the planar gradient coils often suffer from their poor magnetic field linearity. Scaling the spatial frequencies of the current density function designed by the magnetic energy minimization, magnetic field linearity of the planar gradient coils can be greatly improved with small sacrifice of gradient coil inductance. We have found that the figure of merit of the planar gradient coils, defined by the gradient strength divided by the linearity error and the inductance, can be improved by proposed technique.

Keywords