A Study on the Optimal Design for the reconstruction Filter in Single Photon Emission Computed Tomography (SPECT)

단일광자방출 전산화 단층촬영상에서 재구성 필터의 최적설계에 관한 연구

  • 김정희 (포항공과대학교 정보통신연구소) ;
  • 김광익 (포항공과대학교 수학과)
  • Published : 1997.06.01

Abstract

This paper presents an optimal design for the SPECT reconstruction filter, based on a physical limit of SPECT lesion detection capability. To increase the performance of the filter on lesion detectability, the filter design was focused on increasing the local SyW ratio of a threshold lesion, that was determined by minimum detectable lesion size (MDU) from SPECT lesion detectabllity contrast-detail curve. The proposed filter showed flexible window characteristics of resolution recovery and noise smoothing for MDLSs in the resolution-limited and photon-limited regions, respectively, compennting for the relative impact of the main limitation factors on threshold detectability. The simulated results showed good adaptability of the proposed filter to the changes in physical parameters of photon counts, object contrast, and detector system resolution.

Keywords

References

  1. IEEE Trans. Nucl Sci. v.NS-27 Cubic splines for filter design in Computed Tomography S. C. Huang;M. E. Phelps;E. J. Hoffman;D. E. Kuhl
  2. IEEE Trans. Nucl. Sci. v.24 A generalized |W|-filter for 3-D reconstruction Y. S. Kwoh;I.S. Reed;T. K. Trunong
  3. J. Nucl. Med. v.24 Fast count-dependent digital filtering of nuclear medicine images:concise communication M. A. King;P. W. Doherty;R. B. Schwinger;D. A. Jacobs. R. E. Kidder;T. R. Miller
  4. J. Nucl. Med. v.25 Two-dimensional filtering of SPECT images using the Metz and Wiener filters M. A. King;R. B. Schwinger;P. W. Doherty;B.C. Penney
  5. IEEE Trans. on Med. Imaging v.11 Prereconstruction restoration of myocardial Single Photon Emission Computed Tomography images D. Boufelfel;R. M. Rangayyan;L. J. Hahn;R. Kloiber
  6. IEEE Trans. on Biomedical Engineering v.BME-28 The user of filtering methods to compensate for constant attenuation in Single-Photon Emission Computed Tomography G. T. Gullberg;T. F. Budinger
  7. IEEE Trans. Nucl. Sci. v.NS-29 Image noise, resolution and lesion detectability in Single Photon Emission CT C. B. Lim;K. S. HJan;E. G. Hawman;R. J. Jaszczak
  8. Lecture Notes in Medical Informatics: Selected Topics in Image Science Single Photon Emission Computed Tomography: Potentials and Limitations C. B. Lim;S. Gottschalk
  9. IEEE proceedings of the 1996 Medical Imaging Conference A filter design for optimization of lesion detection in SPECT J. H. Kim;K. I. Kim;C. Kwark
  10. Phys. Med. Bio. v.30 Effect of resolution improvement on required count density in ECT imaging:a computer simulation G. Muehllehner