The Calculation of Response Matrix of 2-Dimensional Radiation Monitoring System Using EGS4 Simulation

EGS4 simulation을 이용한 2차원 방사선준위 분포측정 시스템의 Response Matrix 계산

  • Published : 1997.06.30

Abstract

In this study an EGS4 simulation code was used to calculate real energy spectrum from measured ${\gamma}$-ray energy spectrum obtained using 2-dimensional radiation monitoring system. As a result, the $39{\times}39$ response matrix was calculated the energy range of 0.1 to 2 MeV which energy interval of 50 keV The real energy spectrum for Co-60 radioisotope was calculated using inverse of response matrix. It was confirmed that the calculated response matrix was useful to the analysis of the measured energy spectrum for the radiation monitoring system.

2차원 방사선 측정시스템에서 측정된 감마선의 에너지 스펙트럼으로부터 실제 에너지 스펙트럼을 얻기 위하여 CsI(T1) 섬광계수기에 대한 EGS4 모의실험을 실시하였다. $0.1{\sim}2$ MeV의 감마선에 대하여 50 keV 간격으로 모의실험을 실시하여 $39{\times}39$ response matrix를 계산하였고 이에 대한 역행렬을 구하였다. 이를 이용하여 실제 측정된 Co-60 에너지 스펙트럼에 대하여 unfolding을 실시하였다. 그 결과 계산된 response matrix가 실제 스펙트럼 분석에 유용한 것으로 판단되었다.