Experiment for Clinical Application with Photodiode
Photodiode를 사용한 측정기의 임상응용을 위한 실험
- Kim, You-Hyun (Dept. of Radiologic Technology, Junior college of Allied Health Sciences, Korea University) ;
- Kwon, Soo-Il (Dept. of Physics, Kyonggi University) ;
- Huh, Joon (Institute of Radiological Technokogy, Dong-A X-ray Co., L.T.D.)
- Published : 1996.12.20
Abstract
Studies were conducted to determine the clinical application of photodiode. We compared photodiode with ion-chamber as to change of tube potential, tube current, mAs and measured decreasing rate of penetration dose. When tube potential was changed from 60 kVp to 120 kVp, output of photodiode and ion-chamber were changed from 0.4 to 1.625, and 1.018 to 4.268, respectively. This was a good agreement to theory that