A STUDY ON THE ABUTMENT SCREW LOOSENINE OF DENTAL IMPLANTS

치과 임플랜트 지대나사의 풀림 현상에 관한 연구

  • Kang, Yun-Mo (Dept. of Prosthodontics, College of Dentistry, Dankook University) ;
  • Lim, Ju-Hwan (Dept. of Prosthodontics, College of Dentistry, Dankook University) ;
  • Cho, In-Ho (Dept. of Prosthodontics, College of Dentistry, Dankook University)
  • 강윤모 (단국대학교 치과대학 보철학교실) ;
  • 임주환 (단국대학교 치과대학 보철학교실) ;
  • 조인호 (단국대학교 치과대학 보철학교실)
  • Published : 1996.04.01

Abstract

Abutment screw loosening of dental implant is a major problem in the treatment of dental implant. The purpose of this study was to examine the effect on the prevention of screw loosening according to tightening method and using of washer. In the first experiment, using a implant fixture and precisely fabricated temporary cylinder, the loosening torque was measured by tightening the screw 5 times repeatedly. In the second experiment, the fabricated crown and implant fixture was tightened by the screw that is attatched washer and then the loosening torque was measured after loading it 1800, 12600 times under the force of 150N. The results were as follows. 1. The highest loosening torque appeared after tightening 3 times repeatedly(p<0.01). 2. When the washer was used, the loosening torque was increased significantly under the loading 1800 times(p<0.01), 3. When the washer was used, the loosening torque was increaed under the loading 12600 times, but it is statistically not significant(p<0.01). The first experiment results mean that 3 times repeated tightening screw is most effective clinically. The second results mean that washer is effective in prevention screw loosening, but the effect is decreased with increasing load.

Keywords