Development of Cyanide Sensor Using S. cerevisiae

S. cerevisiae를 이용한 시안센서의 개발

  • 김종민 (동아대학교 화학공학과) ;
  • 이현우 (동아대학교 화학공학과 동아대학교 화학공학과 동아대학교 화학공학과 한국과학기술원 화학공학과)
  • Published : 1996.12.01

Abstract

Two types of cyanide sensor system were constructed for monitoring river water simply. One is membrane type flow system composed of S. cerevisiae coated membrane and oxygen electrode, the other is reactor type flow system composed of S. cevevisiae immobilized reactor and two oxygen electrode. S. cerevisiae and oxygen electrode were used for detecting cyanide compounds. These sensors are based on the inhibition of cyanides on the respiration activity of S. cerevisiae. Membrane type sensor system could detect cyanides ranging 0.10∼1.00ppm, and the respiration activity of S. cerevisiae continued for a day. Reactor type sensor system composed of immobilized chitopearl HP-5020 with S. cerevisiae could detect cyanides in the range of 0.10∼1.00ppm, and maintained 90% respiration activity for sixteen days. It was found that the optimum reactor size exist for the reactor type biosensor when respirating activity inhibition is used for detection of cyanides.

간단한 조작으로 시안화물을 신속하게 계측할 수 있는 센서의 개발을 목적으로 하여 미생물과 산소전극을 이용한 시안센서의 개발을 연구하였다. 산소전극에 S. cerevisiae를 흡착 고정화한 막을 부착하여 막형 시안센서를 제작하고 시료용액중의 시 안이온농도를 O~1.00ppm이 되게 시안을 첨가하면 서 전류값 변화를 살펴본 결과 시안농도 0.10~1.00ppm의 범위에서 응답이 확인되었으며 고정화 직후의 호흡활성은 약 하루정도 유지되었다. S. ceremSlae를 고정화한 키토-펼 HP-5020을 사용하여 반응기형 시안센서를 제작하고 막형 시안센서와 동일하게 시안이온농도 O~1.00ppm 용액의 전 류값 변화를 살펴본 결과 0.10~ 1.00ppm의 범위에서 응답이 확인되었으며 고정화 직후 90%의 호흡활 성이 16일간 유지되었다. 반응기형 시안샌서의 반응기 크기의 영향을 검토 한 결과, 반응기의 최적 체적이 존재한다는 사실을 알 수 있었다.