Journal of the Korean Institute of Telematics and Electronics B (전자공학회논문지B)
- Volume 33B Issue 12
- /
- Pages.113-121
- /
- 1996
- /
- 1016-135X(pISSN)
A 37.5kV A Class Power Supply for Diagnostic X -ray by High-Frequency Resonant Inverter Type
37.5kVA급 고주파 공진형 인버터식 진단 X선 전원장치
Abstract
This paper presents a 37.5kVA (125kVp, 300mA) high frequency inverter type power supply for diagonostic X-ray generator with two HFHV (high-frequency high-voltage) isolating transformers. The output regulations, X-ray tuve voltage and tube current, are obtained by two high frequency inverters with the each other feedback PI controller and operating time switch for selecting shooting time and synchronizing two inverters. The operating modes and design consideration of the proposed power supply are given. Issues in the design of two HFHV isolating transformersfors high volage genrating and filament heating are discussed. Experimental results are presented to verify the performance of the designed power supply for varying load conditions. The proposed apparatus has several advnatages, e.g., the fast rising time of tube voltae, accuracy and reduced component size etc.
Keywords