정상유동에서 유동형 단엽폴리머 인공판막의 수력학적 성능평가

Hydrodynamic Investigation of a Floating-type Monoleaflet Polymer Valve under Steady Flow Condition

  • 김준우 (전북대학교 공과대학 기계공학부) ;
  • 박복춘 (전북대학교 공과대학 기계공학부, 서울대학교 의과대학 의공학과)
  • 발행 : 1996.03.01

초록

An experimental investigation was performed under steady flow condition to assess hydrodynamic performance of floating-type monoleaflet polymer valves (MLPV) withdifferent leaflet thickness. The St. Jude Medical valve (SJMV) was also used for comparison test. Pressure drops of MLPVS are larger than those for other types of polymer valves and mechanical valves. Furthermore, the thicker is the leaflet thickness of the polymer valve, the larger are the corresponding pressure drop. The velocity profiles for MLPs reveal a large reversed flow region downward to the valve position. The maximum wall shear stresses of MLPVS at a flow rate of $30{\ell}$/min are in the range 50-130 dyn/$cm^2$, and the corresponding maximum Reynolds shear stresses are in the range of 100-500 dyn/$cm^2$, respectively, which are beyond the allowable limit clinically. In contrast, floating-type monoleaflet polymer valves show better hydrodynamic performance in leakage volume. From the designing point of view, it may be concluded that the optimum thickness of leaflet for better hydrodynamic performance is one of the Important parameters.

키워드

참고문헌

  1. Journal of the Royal Sosiety of Medicine v.76 Twenty-five years of heart valve substitutes : a review M.M. Black;P.J. Drury;W.B. Tindale
  2. CVR&R Artificial heart valves : past performance and future prospects Martin M.;Thomas C.;Philip D.;Patricia V.
  3. J. applied Physiol v.14 no.6 Polyurethane artificial heart valve in animals Akutsu T;Dryer B;Kolff W.J.
  4. Trans. Am. Soc. Artif. Intern. Organs v.24 A single-leaflet mitral valve : its design, development and evaluation in long-term animal trials W.G. Williams;B.J. Belhouse;F.H. Belhouse;W.S. Haworth;G. Kent;R.W.H. Lewis
  5. Trans. Int. Soc. Art. Organs v.2 Design and in-vitro evaluation of a new shell-type artificial heart valve H. Reul;H. Muller;W. Tillman
  6. Trans Am Soc Artif Intern Organs v.XXXV In-vitro and In-vivo Evaluation of a Jellyfish Valve for Practical Use K. Imachi
  7. The development of a prostheic haert valve utilizing a rigid housing and a flexible butterfly-wing leaflet V.L. Gott;R.L. Daggett;R.J. Botham;D.E. Koepke;W.C. Zarnstorff;W.P. Young
  8. The Int. J. of Art. Organ v.13 Development of a new inflow valve for a 20cc semisoft ventricle : preliminary results E.G. Wijsmuller;L.S. Yu;N.D. Bishop;W.G. Kolff
  9. 대한의공학회지 v.13 no.3 이엽식 고분자 판막의 수력학적 특성평가 및 동물실험 김철생;박복춘;서수원 등
  10. KSME J v.7 no.3 In vitro Hydrodynamic Evaluation of Prosthetic Polymer Heart Valves in Steady Flow B.C. Pak;B.S. Cho;B.J. Baek;B.G. Min
  11. J. of Thor. Card. surg v.49 A late evaluation of flexible teflon prosthesis utilized for total aortic valve replacement : postoperative clinical, hemodynamic and pathological assessment Brauwald N.S.
  12. Trans, Amer. Soc. Artifi. Int. Org v.26 Development of seamless trileaflet valves Russel F.B.
  13. Amer. Soc. Art. Int. Org. Abstracts v.11 A polyurethane trileaflet cardiac valve prosthesis : in vivo and in vitro stedies Wisman C.B.
  14. J. of Clinical Engineering v.16 In-vitro engineering testing of prosthetic cardiac valves Yosr A.M. Haggag
  15. Rdfraction correction for LDA measurements in flows with cerved optical boundaries v.VIII no.Issue2 A.F. Bicen
  16. American J. Cardiology v.38 Choosing a Substitute Cardiac Valve : Type, Size, Surgeon Roberts;W.C.
  17. J. Biomech. Mat. Res. v.8 Adhesion of Red Cells to Foreign Surfaces in the Presence of Flow Mohandas;N;Hochmuth, R.M.;Spaeth, E.E.
  18. Circulation Res v.22 Acute Vascular Endothelial Changes Associated with Increased Blood Velocity Gradient Fry, D.L.
  19. Proc.6th Int. Conf. on Bio. Eng. Hemodynamics of Tilting Disc-Type Prosthetic Heart Valves Modi, V.J.;Bishop, W.F.
  20. J. Biomechanics v.16 Laser Anemometry Measurements of Pulsatile Flow Past Aortic Valve Prostheses Chandran, K.B.;Cabell, G.N.;Khalighi, B.;Chen, C.J.
  21. Trans. Am. Soc. Artif. Internal Organs v.22 Shear Induced Aggregation and Lysis of Platelets Hung, T.C.;Hochmuth, R.M.;Joist, J.H.;Sutera, S.P.
  22. J. Biomechanics v.28 no.4 A Numerical Analysis of the Back Flow between the Leaflets of a St. Jude Medical Cardiac Valve Prosthesis Reif, T.H.
  23. ASAIO Trans v.37 Mean Velocities and Reynolds Stresses within Regurgitant Jets Produced by Tilting Disc Valves Baldwin, J.T.;Tarbell, J.M.;Deutch S.;Gesselowitz D.B.