Dose Distribution and Characterization for Radiation Fields of Multileaf Collimateor System
방사선 입체조형치료용 다엽콜리메이터의 특성과 조직내 선량분포 측정
- Chu, Sung-Sil (Department of Radiation Oncology, College of Medicine, Yonsei University) ;
- Kim, Gwi-Eon (Department of Radiation Oncology, College of Medicine, Yonsei University)
- Published : 1996.03.01
Abstract
Purpose : Multileaf collimator(MLC) is very suitable tool for conformal radio-therapy and commissioning measurements for a multileaf collimator installed on a dual energy accelerator with 6 and 10MV photons are required, For modeling the collimator with treament planning software, detailed dosimetric characterization of the multileaf collimator including the penumbra width, leaf transmission between leaf leakage and localization of the leaf ends and sides is an essential requirement. materials and Methods : Measurement of characteristic data of the MLC with 26 pair block leaves installed on CLINAC 2100C linear accelerator was performed. Low sensitive radiographic film(X-omatV) was used for the penumbra measurement and separate experiments using radiographic film and thermoluminescent dosimeters were performed to verify the dose distribution, Measured films were analized with a photodensitometer of WP700i scanner. Results : For 6 & 10 MV x-ray energies, approximately
목적 : 다엽콜리메이터를 이용하여 부정형의 치료부위를 편리하고 정확히 차폐하며 삼차원 입체조형치료를 원활히 수행하기 위하여 다엽콜리메이터의 차폐특성을 계측하고 치료계획을 위한 최적방법을 확정한다. 대상 및 방법 : 차폐폭이 1cm이고 길이가 15cm인 52개의 금속차폐엽들이 양방향 각각 26개로 구성되어 컴퓨터계획에 의거 여러모양의 차폐면을 구사할 수 있는 다엽콜리메이터(Multileaf collimator)를 대상으로 방사선 치료에 요구되는 고에너지 방사선의 출력선량율, 차폐 및 투과효과, 반음영, 조직내 선량분포등을 전리함측정기와 열형광측정기 및 필름 선량측정방법을 이용하여 계측하였다. 다엽콜리메이터의 계단식 반음영 측정은 고체팬텀과 비디오밀도측정기를 이용하여 실효반음영(Effective penumbra)과 일반적반음영(General penumbra)으로 구분 평가하였다. 결과 : 다엽콜리메이터의 출력선량율은 표준차폐조리개에 비교하여 출력 선량율이