The Measurement of Dose Distribution in the Presence of Air Cavity and Underdosing Effect Result from Lack of Electronic Equilibrium

조사면내 공동의 존재에 따른 선량분포의 변화측정

  • Jo, Jeong-Hui (Dept. Radiation Oncology, Yonsei University College of Medicine) ;
  • Bang, Dong-Wan (Dept. Radiation Oncology, Yonsei University College of Medicine) ;
  • Park, Jae-Il (Dept. Radiation Oncology, Yonsei University College of Medicine)
  • 조정희 (연세의료원 암센터 방사선종양학과) ;
  • 방동완 (연세의료원 암센터 방사선종양학과) ;
  • 박재일 (연세의료원 암센터 방사선종양학과)
  • Published : 1996.11.01

Abstract

When high energy photon beam is incident upon an air cavity interface the effect of ionization build-up observed . This phenomenon is resulting from the surface layers of the lesions are significant deficiency of electrons reaching the layers because of the replacement, of solid scattering material by the air cavity, that is lack of electronic equilibrium. Measurement have been made in an acrylic phantom with a parallel plate chamber and high energy photon beams, CO-60, 4MV, 6MV and 10MV X-rays have been investigated. The result of our study show that a significant effect was measured and was determined to be very dependent on field size, air cavity dimension and photon energy. The reductions were much larger for 10MV beam, underdosage at the interface was 12, 12.2, 16.9 and $20.6\%$ for the CO-60, 4MV, 6MV and 10MV, respectively. It was found that this non-equilibrium effect at the interface is more severe for the higher energy beams than that of lower energy beams and the larger cavity dimensions the larger beam reductions occur. This problem is of clinical concern when lesions such as carcinoma beyond air cavities are irradiated, such as larynx, glottic and the patients with maxillectomy and ethmoidectomy and so forth.

Keywords