The Journal of the Acoustical Society of Korea (한국음향학회지)
- Volume 14 Issue 2
- /
- Pages.30-39
- /
- 1995
- /
- 1225-4428(pISSN)
- /
- 2287-3775(eISSN)
A study on the DSP Analysis for the CAT application
CAT 응용을 위한 신호처리 분석에 관한 연구
Abstract
In this paper, study on implementation of FFT analyzer applied to CAT, A/D conversion module, DSP module and VXIbus interface module are implemented in hardware and calculation program and control software are implemented in DSP module and VXIbus interface module, respectively. The control of the modules using PC is realized in software. The real time bandwidth of the FFT analyzing device is 100KHz. At sampling rate of 200KHz and with 2048 point FFT, the result of applying sine, triangular and rectangular wave of 20KHz to FFT analyzing device is compared with the FFT analyzed results of Hewlett-Packard 3562A dynamic output range of -40dBV- +30dBV, correct results are obtained and results of applying 10KHz, 20KHz and 50KHz input are compared and the correct values are obtained.
본 논문에서는 A/D변환 모듈, 신호처리 모듈, VXI 인터페이스 모듈의 하드웨어를 구성하였고 신호처리 모듈과 VXI 인터페이스 모듈 각각에 연산 및 제어용 소프트웨어를 구현하였다. FFT 분석 디바이스의 실시간 대역폭은 100KHZ라고 말할 수 있다. 샘플링을 200KHz에서 point수 2048의 연산 파라미터 하였을 때 20KHz 입력신호의 정현파, 삼각파 및 구형파에 대한 결과를 Hewlett-Packard 3562A Dynamic Signal Analyzer낱 이용한 FFT 분석결과와 각각 비교하였다 주파수 SPAN을 0-100KHz까지로 하고 출력범위를 -40dBV 에서 30dBV 까지로 하였을 때의 결과로서 매우 정확하게 분석됨을 알 수 있었으며, 입력주파수를 10KHz, 20KHz, 50KHz로 각각 주었을 때의 결과를 비교했을 때 원하는 결과를 얻을 수 있다.
Keywords