Determination of Ammonia-N in Environmental Water by Air-segmented FIA

  • Feng, Yong-Lai (Department of Chemistry, Nanjing University) ;
  • Tian, Li-Ching (Centre of Materials Analysis, Nanjing University) ;
  • Wang, Wei (Centre of Materials Analysis, Nanjing University) ;
  • Wu, Qiao-Znen (Centre of Materials Analysis, Nanjing University)
  • Received : 1995.08.20
  • Published : 1995.12.25

Abstract

A method for determination of ammonia-N in environmental water by air-segmented flow injection analysis using the colour reaction of phenol and sodium hypochlorite with ammonia was described in this paper. When the reaction temperature is $70^{\circ}C$ and the reaction residence time is 5 minutes, a sample frequency of $60-90h^{-1}$ can be achieved. The detection limit and relative deviation are $0.05mg.ml^{-1}$ and 4%, respectively. The recovery of this method is 96 - 110%, and the determination results of the method are greatly agreement with standard colorimetric method.