Abstract

Ion-selective poly(vinyl chloride) membrane electrodes for the determination of the anticholinergic drug alverine in pure substances is described. Ion-pair complex of alverine with anionic counter ion acid red 97 is investigated as electroactive compound for the electrode membrane. On the optimal composition of membrane, its slope was 55.35mV/decade, relative standard deviation was 0.61 and lower limit of linear range was $1.0{\times}10^{-5}M$. The observed potentiometric selectivity coefficient was -2.625 toward methylephedrine and -2.216 toward histidine. We can got the stable potential at pH 7.0~4.0. Response time was 20~30 seconds for solutions${\leq}10^{-4}M$, about 10 seconds for solutions ${\geq}10^{-4}M$ of these compounds.

Alverine에 대한 anionic counter ion으로서 acid red 97의 이용 가능성을 살펴보았다. PVC와 가소제의 비율이 같을 때 적절한 감응과 긴 수명을 나타내었으며, 검량선은 $1.0{\times}10^{-5}M$까지 직선성이었고 그 감응 기울기는 55.35mV/dec., 변동계수는 0.61이었다. Methylephedrine에 대하여는 -2.626, histidine에 대하여는 -2.216등의 선택계수를 보였고 pH 7.0~4.0에서 안정된 전위값을 얻을 수 있었다. $10^{-4}M$ 이하에서 20~30초 이내, 그 이상의 농도에서는 약 10초 이내에서 안정된 전위값을 얻을 수 있었다.