$^{99m}Tc-MAG_3$ 제거지수를 이용한 이식신장의 기능평가

$^{99m}Tc-MAG_3$ Elimination Index on Normal Functioning Transplanted Kidney

  • 전우진 (계명대학교 의과대학 진단방사선과학교실) ;
  • 김주헌 (계명대학교 의과대학 진단방사선과학교실) ;
  • 박미옥 (계명대학교 의과대학 진단방사선과학교실) ;
  • 이희정 (계명대학교 의과대학 진단방사선과학교실) ;
  • 현정애 (계명대학교 의과대학 진단방사선과학교실) ;
  • 전석길 (계명대학교 의과대학 진단방사선과학교실)
  • Jeon, Woo-Jin (Department of Diagnostic Radiology, School of Medicine, Keimyung University) ;
  • Kim, Ju-Heon (Department of Diagnostic Radiology, School of Medicine, Keimyung University) ;
  • Park, Mi-Ok (Department of Diagnostic Radiology, School of Medicine, Keimyung University) ;
  • Lee, Hee-Jung (Department of Diagnostic Radiology, School of Medicine, Keimyung University) ;
  • Hyun, Jung-Ae (Department of Diagnostic Radiology, School of Medicine, Keimyung University) ;
  • Zeon, Seok-Kil (Department of Diagnostic Radiology, School of Medicine, Keimyung University)
  • 발행 : 1995.03.31

초록

정상기능 이식신장 50예와 비정상기능 이식신장 7예의 $^{99m}Tc-MAG_3$를 이용한 신장스캔에서 EI를 구하여 이식신장의 배설기능을 정량적으로 측정하였다. 이식신장의 $^{99m}Tc-MAG_3$ 신장기능곡선에서 정상기능 이식신장의 평균 EI는 $2.21{\pm}0.51$이었으며, 95% 신뢰구간은 2.01-2.87이었다. 비정상기능 이식신장에서는 평균 EI가 $0.74{\pm}0.18$로 1이하의 수치를 보여, 이식신장의 간질에 임파구의 침윤이나 세뇨관의 괴사등에 의해 관류와 배설기능의 장애를 초래하여 신실질과 수집관내에 방사능 동위원소의 저류가 나타났음을 뜻한다. 그리고 EI, $T_{max}$, $T_{1/2}$, BUN, 혈청 Creatinine의 정상기능 이식신장군과 비정상기능 이식신장군의 비교에서 통계학적으로 유의한 차이가 있었다(p<0.0001). 그러므로 이식 후 EI의 측정으로 이식신장의 기능상태를 알 수 있고 거부반응 및 신세뇨관 괴사 등의 합병증을 조기에 찾아내는 지표로 사용할 수 있을 것으로 사료된다.

Purpose : We analysed $^{99m}Tc-MAG_3$ renal scans to evaluate renal function of transplanted kidney and to detect various renal transplant complications, measuring the ratio of renal radioactivity at three minutes to that at 20 minutes(elimination index). Material and Methods : The fifty seven renal transplantation recipients were studied. There were 50 normal functioning transplanted kidneys as group I and 7 abnormal function-ing transplanted kidney, including 5 cases of acute renal rejection, 2 cases of acute tubular necrosis as group IIl. The protocol consisted of: (1) $^{99m}Tc-MAG_3$ 740MBq injection intravenously : (2) sequential imaging for 2min(60two-second images) followed by 30min(30 sixty-second images) : (3) drawing of region of interest(ROI) on renal imaging; (4) time-activity corves were generated from renal ROI after background subtraction, and time of maximum activity($T_{max}$) and half time of maximal peak radioactivity($T_{1/2}$) were produced in the renogram curve. (5) EI through Bischof-Delaloye method as determined on the renogram curve. Results : Normal group( I ) shows mean EI of 2.21(95.0% Confidence limit of 2.01-2.87), $T_{max}$ of 154 sec, $T_{1/2}$ of 1,139 sec. Abnormal group(II) shows mean EI of 0.74, $T_{max}$ of 1,466 sec, $T_{1/2}$ of 19,224 sec. The EI, $T_{max}$, $T_{1/2}$, BUN and serum creatinine values are significantly different between normal group(I) and abnormal group(II) (p<0.0001). Conclusion : By measuring EI with $^{99m}Tc-MAG_3$, renal function of transplanted kidney could be easily evaluated and various complications could be detected early.