Construction of the NQR Gaussmeter using Computer Interface

Computer Interface를 이용한 핵 사중극 공명 Gaussmeter의 제작

  • Published : 1995.04.01

Abstract

When the low magnetic field is applied to the synunetry axis of the axial synunetric electric field gradient tensor of the quadrupolar nucleus in the crystal with nuclear spin, I=1, two split resonance frequencies are observed by the Zeeman effect on the nuclear quadrupole resonance. The applied magnetic field is proportional to the difference of the two resonance frequencies and this proportionality constant for the $^{14}N$ of ${(CH_{2})}_{6}N_{4}$ single crystal is 0.16 mT/kHz. The NQR spectrometer is interfaced with a personal computer from which the resonance signals are displayed and the value of magnetic field is obtained directly from the difference of the two resonance frequeocies. The lowestest measured magnetic field was 0.20 mT using this NQR technique.

전기장 기울기 텐서가 축대칭인 결정에 대칭인 방향으로 자기장을 가하면 핵사중극공명에 미치는 Zeeman 효과에 의하여 두 개의 분리된 공명진동수가 측정된다. 이 때 가해준 자기장과 두 공명진동수 간의 차이는 서로 비례하며, ${(CH_{2})}_{6}N_{4}$ 단결정의 $^{14}N$ 원자핵에 대한 비례상수는 0.16 mT/kHz 이다. 핵 사중극 공명장치 와 개인용 컴퓨터를 interface 하여 핵 사중극 공명신호의 모습을 모니터에 표시하여 측정되는 두 공명신호의 진동수 차이로 자기장을 직접 읽을 수 있도록 하였다. 여기로부터 핵 사중극 공명 Gaussmeter를 이용하여 실험적으로 측정한 최저 자기장은 0.20 mT 였다.