DESIGN OF A RISE TIME DISCRIMINATION CIRCUIT FOR X-RAY PROPORTIONAL COUNTER

X-선 비례계수관용 상승시간 선별회로 제작

  • 남욱원 (한국표준과학연구원 천문대) ;
  • 최철성 (한국표준과학연구원 천문대)
  • Published : 1995.06.01

Abstract

It is possible to distinguish X-ray events from non X-ray events in proportional conters using the method of rise time discrimination (RTD). In order to subtract non X-ray background, we have developed a simple RTD circuit which will be applied to the proportional counter planned for a sounding rocket experiment. The entire circuit consists of two parts ; the rise time measurement circuit and the time to amplitude conversion circuit which includes the self-calibration mode. From the test with X-ray detecting system, we obtained that the background can be rejected more than 80% in the energy band 2∼12 keV. However we confirmed that the RDT method is not proper to be used for the energy range above 12kV.

비례계수관에서 생성되는 신호의 상승시간을 비교하여 X-선과 비 X-선 신호를 구분하는것이 가능하다. 과학로케트 탑재형 비례계수관에 응용하여 백그라운드를 제거하기 위한 간단한 상승시간 선별회로를 개발하였다. 제작된 전체 회로는 상승시간 측정회로, 측정된 시간을 진폭으로 변환시켜주는 시간-진폭 변환회로로 구성되며, 시산 대 진폭을 검교정할 수 있는 회로를 부가하였다. 제작된 회로를 X-선 검출 시스템에 적용한 결과 2~12 keV에너지 범위에서 전반적인 백그라운드 제거율을 80% 이상 얻을 수 있었으며, 약 12 keV 이상의 높은 에너지내에서는 상승시간 선별법을 사용하기 어려운 제한성을 확인하였다.

Keywords