Hemodynamic Evaluation of Monoleaflet Polymer Valve

단엽식 고분자판막의 혈역학적 성능평가

  • 김상현 (연세대학교 의과대학 심혈관연구소) ;
  • 장병철 (연세대학교 의과대학 흉부외과학교실, 동아대학교 기계공학과, 펜실베니아주립대학교 의공학과)
  • Published : 1995.03.01

Abstract

We have developed a monoleaflet polymer valve as an inexpensive and viable alternative, especially for short-term use in the ventricular assist device or total artificial heart. The frame and leaflet of the polymer valve were made from polyurethane. To evaluate the hemodynamic performance of the polymer valve a comparative study of flow dynamics past a polymer valve and a St. Jude Medicals prosthetic valve under physiological pulsatile flow conditions in vitro was made. Comparisons between the valves were made on the transvalvular pressure drop, regurgitation volume and maximum valve opening area. The polymer valve showed smaller regurgitation volllme and transvalvular pressure drop compared to the mechanical valve at higher heart rate. The results showed that the functional characteristics of the polymer valve compared favorably with those of the mechanical valve at higher heart rate.

Keywords

References

  1. J. appl. Physiol. v.14 Polyurethane artificial heart valves in animals Akutsu,T.;Dreyer,B.;Kolff,W.J.
  2. ASAIO Trans. v.28 A polyurethane trileaflet cardiac valve prosthesis in vitro and n vitvo studies Wisman,C.B.;Pierce,W.S.;Donachy,J.H.;Pae,W.E.;Myers,J.L.;Prophet,G.A.
  3. Polyurethanes in Biomedical Engineering Ⅱ The Helmholtz Institute trileaflet polyurethane heart valve prosthesis: design manufacturing and first in vitro and in vitro results Harold,M.;Lo,H.B.;Reul,H.;Machter,H.;Tauchi,K.;Giersiepen,M.;Birkle,G.;Hollweg,G.;Rau,G.;Messmer,B.J.;Planck,H.(ed. et al)
  4. J. Thorac. Cardiovasc. Surg. v.94 Evaluation of explanted polyurethane trileaflet cardiac value prosthesis Hilbert,S.L.;Jones,M.
  5. ASAIO Trans. v.35 New polyurethane valves in new soft artificial hearts Yu,L.S.;Yuan,B.;Bishop,D.;Topaz,S.;Griensven,J.V.;Hofma,S.;Swier,P.;Kinkmann,J.;Kolff,J.;Kolff,W.J.
  6. ASAIO Trans v.V57 no.3 Advances in design principle and fluid dynamics of a flexible polymeric heart valve Jansen,J.;Willeke,S.;Reiners,B.;Harbott,H.;Reul,H.;Lo,H.B.;Dabritz,S.;Rosenbaum,C.;Bitter,A.;Ziehe,K.;Rau,G.;Messmer,B.J.
  7. ASAIO Trans. v.24 A single-leaflet mitral valve: its design, development and evaluation in long-term animal trials Williams,W.G.;Bellhouse,B.J.;Bellhouse,F.H.;Haworth,W.S.;Kent,G.;Lewis,R.W.H.
  8. ASAIO Trans. v.34 A newly designed jellyfish valve for an artificial heart blood pump Imachi,K.;Fujimasa,I.;Mabuchi,K.;Chinzei,T.;Abe,Y.;Maeda,K.;Imanishi,K.;Kouro,A.;Ono,T.;Atsumi,K.
  9. The Int J of Art Organ v.13 Development of a new inflow valve for a 20cc semisoft ventricle:preliminary results Wijsmuller,E.G.;Yu,L.S.;Yuan,B.;Bishop,N.D.;Kolff,W.J.
  10. 의공학회지 v.13 이엽식 고분자판막의 수력학적 특성평가 및 동물실험 김철생;박복춘;서수원;한동근;이규택;최진욱(등)
  11. 대한의용 생체공학회 춘계학술대회 논문집 v.15 no.1 인공심장 및 심실보조장치용 고분자 인조판막의 개발 서수원;J.E.v.d.;Wetering;박영주;박성근;김인영;민병구
  12. Art. Org. v.15 vitro hemodynamic analysis of flexible artificial ventricles Chandran,K.B.;Lee,C.S.;Shipkowitz,T.;Chen,L.D.;Yu,L.S.;Wurzel,D.
  13. J. of biomechanical Engineering v.114 Numerical simulation of steady flow in a two-dimensional total artificial heart model Kim,S.H.;Chandran,K.B.;Chen,C.J.
  14. 대한의용 생체공학회 추계학술대회 논문집 v.15 no.1 응력 및 변위를 최소화 하기위한 단엽식고분자 판막의 설계 한근조;김상현
  15. Am. Heart J v.41 Hydraulic formula for the calculation of the area of strentic mitral valve. other cardiac valves and central circulatory shunts Gorlin,R.;Golin,S.G.
  16. J. Thoracic & Cardiovas. Surg. v.76 no.6 vitro hydrodynamic comparison of mitral valve prosthessis at high flow rates Gabbay,S.;McQueen,D.M.;Yellin,E.L.;Frater,R.W.M.