고선량율 근접치료기의 선원교정과 치료실주변 방사선량 측정

Calibration and Radiation Survey of High Dose Rate Remote Afterloading System

  • 이정옥 (원광대학교 의과대학 치료방사선과학교실) ;
  • 강정구 (전주 예수병원 치료방사선과) ;
  • 문성록 (원광대학교 의과대학 치료방사선과학교실)
  • Lee, Jeon-Gok (Department of radiation Oncology, Wonkwang University, School of Medicine) ;
  • Kang, Jeong-Ku (Department of Radiation Oncology, Chonju Presbyterian Medical Center) ;
  • Moon, Sun-Rock (Department of radiation Oncology, Wonkwang University, School of Medicine)
  • 발행 : 1995.03.01

초록

High Dose Rate Remote Afterloading system was installed at Wonkwang University Hospital in January 1994. In this report, the calibration of a Gammamed 12-i High Dose Rate Remote Afterloading system and the radiation survey around the facility after design and construct a shieding room are discussed. The radiation survey of the facility indicates that the use of ordinary concrete shielding of existing room will provide adequate shielding. Also, the methodologies for performing source calibration are presented.