A Rapid Method for the Measurement of $^{222}Rn$ in Groundwater and Hot Spring Water using Ultra Low-Level Liquid Scintillation Counter and Pulse Shape Analysis

극저준위 액체섬광계수기와 파형분석법을 이용한 지하수 및 온천수중 $^{222}Rn$의 신속측정법

  • Published : 1995.06.30

Abstract

For the determination of $^{222}Rn$ in water by a very simple and time saving liquid scintillation counting method that does not require any chemical separation, an optimum analytical condition has been investigated. The optimum pulse shape analysis(PSA) level for the measurement of $^{222}Rn$ using LKB 1220 Quantulus liquid scintillation counter was 110 for Optiphase HiSafe3 cocktail and 90 for toluene-based cocktail. The counting efficiencies of $^{222}Rn$ in the window covering u spectra using Optiphase HiSafe3 cocktail were 282.2% for glass vial and 271.6% for Teflon vial, whereas the counting efficiencies in toluene-based cocktail were 262.3% and 247.5% for glass and Teflon vials, respectively. The minimum detectable activity(MDA) in the u window for a 60-min measurement with a Teflon vial using Optiphase HiSafe3 cocktail was $0.30Bq/{\iota}$. The analytical method studied from this work was also applied to the determination of $^{222}Rn$ in some groundwater and hot spring water samples.

계측이 간단하며 화학적 분리가 필요 없는 액체섬광계수법을 이용하여 일부 지하수 및 온천수중 $^{222}Rn$의 최적 분석조건을 검토하였다. $^{222}Rn$의 분석을 위한 최적 파형분석(PSA) 준위는 Optiphase HiSafe3 형광체 (Cocktail)에서 110, 그리고 톨루엔계열 형광체 (Toluene-based cocktail)에서는 90이었다. $^{222}Rn$의 계측효율은 Optiphase HiSafe3 형광체를 이용한 경우 ${\alpha}$영역에서의 계수효율은 유리용기는 282.2%, 테프론용기에서는 271.6%였으며, 톨루엔계열 형광체를 이용한 경우에는 유리용기와 테프론 용기에서 각각 262.3%, 247,5%였다. Optiphase HiSafe3 형광체와 테프론용기를 사용하여 60분간 계측하는 경우 ${\alpha}$-선 피크영역에서 $^{222}Rn$의 검출하한치는 $0.30Bq/{\iota}$였다. 또한, 본 연구에서 검토한 분석법은 일부 지하수 및 온천수중 $^{222}Rn$의 분석에도 응용되었다.

Keywords