Journal of Korean Academy of Fundamentals of Nursing (기본간호학회지)
- Volume 1 Issue 1
- /
- Pages.51-68
- /
- 1994
- /
- 1225-9012(pISSN)
- /
- 2287-1802(eISSN)
Comparative Study About The Indirect Blood Pressure (measured by different Instruments and Methods)
측정도구와 방법에 따른 간접혈압측정치의 비교연구
- Suh Gil-Hee (Dept of Nursing, National Wonju Junior College)
- 서길희 (원주전문대학 간호학과)
- Published : 1994.02.28
Abstract
One aim of this study was to find out the difference of indirect blood pressure between the types of instruments. The difference of indirect blood pressure was measured with mercury sphygomomanometer and electronic blood pressuremeter. Another was to determine the difference of indirect blood pressure according to width of cuff and site of cuff. Indirect blood pressures were measured with the diaphragmatic side of sphygomomanometer. The subjects were 40 female volunteers, whose arm circumferences were from 22 to 34cm. The data were analyzed by SPSS-PC program and pared t-test was used. The results are summarized as follows ; 1. Mercury sphygomomanometer systolic pressure was higher(average 2.20mmhg) than electronic blood pressuremeter. It was statistically significant(p=0.026). 2. The value of Mercury sphygomomanometer diastolic pressure was lower(2.00mmhg) than electronic's. It was statistically significant(p=0.03) 3. In the mercury sphygomomanometer systolic pressure, the value of standard cuff(