Automatic measurement of gage block using digital image processing

디지탈 화상처리를 이용한 게이지 블록의 측정 자동화

  • Ko, Y.U. (Dynamics research Center, Korea Reserch Institute of Standards and Science) ;
  • Suh, H.S. ;
  • Eom, T.B.
  • Published : 1994.10.01

Abstract

We have carried out a research on automatic measurement of gage blocks below 250mm and then improved the accuracy of the system, and reduced the measurement time using image processing technique and computer interface. The accuracy of the system is (230+10330 $L^{2}$)$^{2}$ nm (L:m) with confidence level of 95%. We participated APMP gage block intercomparision-1993/1994. The 1st round intercomparision showed that the measured values of all 5 gage blocks were within 21 nm from average.

Keywords