Time-resolved Fluoroimmunoassay for the Measurement of 17$\beta$-Estradiol using Anti-idiotypic Antibody

Anti-idiotype 항체를 이용한 17$\beta$-Estradiol 측정을 위한 Time-resolved Fluoroimmunoassay

  • 김윤규 (건국대학교 축산대학 동물자원연구소) ;
  • 김창규 (건국대학교 축산대학 동물자원연구소) ;
  • 박성민 (건국대학교 축산대학 동물자원연구소) ;
  • 이치호 (건국대학교 축산대학 동물자원연구소) ;
  • 이원창 (건국대학교 축산대학 동물자원연구소) ;
  • 최영숙 (건국대학교 의과대학부속 민중병원 임상병리과) ;
  • 김종배 (건국대학교 축산대학 동물자원연구소)
  • Published : 1993.02.01

Abstract

A competitive type immunoassay method for 17$\beta$-estradiol(E2) based on the idiotypic anti-idiotypic antibody and time-resolved fluorescence is described. The anti-idiotypic antibody(Ab2) produced to E2 binding site of the primary idiotype antibody (Ab1) was labelled with europium and was allowed to compete with E2 standards or serum sample for the binding sites of Ab1 which was bound to 2nd antibody captured ontothe surface of microtitre plates. Fluorescence measured by time-resolved fluorometer was inversely proportional to the concentration of E2 over the range 5~500pg/well. The sensitivity of the assay was 5pg per well which was compatible with that ofradioimmunoassay using the same Ab1 and 3H-E2 as a tracer. One great advantage of this method described here was to enable antibodies to be labelled instead of haptens, and thus makes it easier to develop sensitive and robust immunoassay systems specially for haptens.