Anisotropy and Dose Equivalents Conversion Factors for the Unmoderated $^{252}Cf$ Source

비감속 $^{252}Cf$ 중성자선원에 대한 비등방성교정인자 및 선량당량환산인자

  • Published : 1993.11.30

Abstract

Form the pure Maxwellian distribution(kT= 1.42MeV), the effects upon calibration factors of encapsulating a $^{252}Cf$ spontaneous fission neutron source were investigated to establish a standard neutron field in the Secondary Standard Dosimetry Laboratory at Korea Atomic Energy Research Institute(KAERI). A Monte Carlo code MCNP was used in simulating the encapsulation SR-Cf-100 and SR-Cf-1273 to be real conditions. The anisotropy(FI) and fluence-to-dose equivalents conversion factors$(H/{\Phi})$ were evaluated and compared with other results. As the results, the FI was determined to be 1.061 at ${\theta}=90^{\circ}$ with ${\pm}0.2%$ statistical error and the $(H/{\Phi})$ was evaluated to be $333.9 [pSv\;cm^2]\;with\;{\pm}0.5%$ statistical error, which is lower by 1.8% than that recommended by the ISO 8529. This means physically that the neutron spectrum of the unmoderated $^{252}Cf$ source in KAERI is a little more softened than that by the ISO.

중성자 측정장비 교정을 위한 표준중성자장을 제작하기 위하여 순수멕스월분포(kt=1.42MeV)로부터 $^{252}Cf$ 자발핵분열 중성자선원의 밀봉이 교정인자에 미치는 영향을 고찰하였다. SR-Cf-100과 SR-Cf-1273 밀봉모형을 실제 제작조건으로 하여 MCNP 코드를 사용하여 몬테카를로 모의를 수행하여, 비등방성교정인자와 선속밀도-대-선량당량 환산인자를 산정하였고, 다른 연구결과와 비교하였다. 결과로서, $FI({\theta}=90^{\circ})$는 1.061(통계오차 : ${\pm}0.2%$), $H/{\Phi}$$333.9(pSv\;cm^2)$ (통계오차 : ${\pm}0.5%$)인 것으로 산정되었다. 이 환산인자$(H/{\Phi})$의 값은 ISO 8529의 권고보다 1.8%가 작은 것인데, 이것은 한국원자력연구소의 비감속 $^{252}Cf$중성자선원의 스펙트럼이 ISO의 것보다 약간 더 연화하다는 물리적 의미를 갖는다.

Keywords